Model Number SC-2408-56
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Impaired Healing (2378)
Date of Event 01/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had an infection at the spinal cord stimulator (scs) leads site due to impaired wound healing.The physician believed that the infection was not procedure or device related.The patient underwent a procedure wherein the leads and ipg were explanted and was doing well postoperatively.The patient was placed on antibiotics and the explanted devices were discarded by the medical facility and will not be returned.
 
Additional Manufacturer Narrative
Additional suspect medical device components involved in the event: product family: scs-linear leads-mri, upn: m365sc2408560, model: sc-2408-56, serial: (b)(6), batch: 7090192, udi: (b)(4).Product family: scs-ipg-r-mri, upn: m365sc12320, model: sc-1232, serial: (b)(6), batch: 798429, udi: (b)(4).Product family: scs-lead fixation-mri, upn: m365sc43190, model: sc-4319, batch: 34856782, udi: (b)(4).
 
Additional Manufacturer Narrative
Investigation results: all explanted devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed that this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: the device was not returned for analysis and the complaint as reported could not be confirmed.Record review revealed no additional information related to the complaint.Therefore, this investigation is unable to determine a probable cause for the complaint and the conclusion code known inherent risk of device will be used.
 
Event or Problem Description
It was reported that the patient had an infection at the spinal cord stimulator (scs) leads site due to impaired wound healing.The physician believed that the infection was not procedure or device related.The patient underwent a procedure wherein the leads and ipg were explanted and was doing well postoperatively.The patient was placed on antibiotics and the explanted devices were discarded by the medical facility and will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVISTA? MRI
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key24440336
Report Number3006630150-2026-01088
Device Sequence Number21387612
Product Code LGW
UDI-Device Identifier08714729904816
UDI-Public(01)08714729904816(17)270727(10)7089048
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSC-2408-56
Device Catalogue NumberSC-2408-56
Device Lot Number7089048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2026
Supplement Date Received by Manufacturer04/07/2026
Initial Report FDA Received Date02/25/2026
Supplement Report FDA Received Date04/30/2026
Date Device Manufactured07/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-