• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA EVITA V800; VENTILATORS, INTENSIVE CARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA EVITA V800; VENTILATORS, INTENSIVE CARE Back to Search Results
Model Number 8422500
Medical Device Problem Codes No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event or Problem Description
It was reported that the device stopped during use.No patient health consequences have been reported.
 
Additional Manufacturer Narrative
The affected device was analyzed via provided information incl.Secured log files.According to the log, it could be identified that the error messages ¿device failure (14)¿ and the subsequent ¿alarm system failed¿ were issued at the reported time.Both alarms resulted from a temporary impairment of the internal hdbaset communication between the ecd and the ventilation unit.In such cases, display functions may be affected and the screen may turn black.In case of an internal failure of the display unit (ecd) and/or an impaired internal hdbaset communication between the ventilation unit and the ecd the alarm message "device failure (14)" will be raised.The display functions and operability of the ecd might be affected in case of an active "device failure (14)" alarm event depending on the technical root cause.If the internal hdbaset communication fails completely, the secondary alarm (piezo speaker) of the ventilation unit will sound.Safety relevant parameters as fio2, minute volume and airway pressure are displayed on the supplemental oled-display wherein the user can observe the ongoing ventilation.The device reacted as specified and needs to be checked according to the proper fitment of system cable and pba syscon ecd.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event or Problem Description
It was reported that the device stopped during use.No patient health consequences have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVITA V800
Common Device Name
VENTILATORS, INTENSIVE CARE
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23558
GM   23558
Manufacturer Contact
moislinger allee 53-55
lübeck 23558
4518822868
MDR Report Key24441184
Report Number9611500-2026-00080
Device Sequence Number10211957
Product Code CBK
UDI-Device Identifier04048675542148
UDI-Public(01)04048675542148(11)210126(93)8422500-04
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K222024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8422500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/11/2026
Supplement Date Received by Manufacturer04/01/2026
Initial Report FDA Received Date02/25/2026
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-