It was reported that, during reprocessing, the sigmoidovideoscope tested positive for >50 colony forming units (cfus) of pseudomonas aeruginosa, 18 colony forming units (cfus) of acinetobacter species, and 2 colony forming units (cfus) of gram-negative rod.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
This report is being supplemented to provide additional information based on the completed investigation's review of the results of third-party testing, the device evaluation and the final investigation.Despite good faith attempts the user cleaning disinfection and sterilization (cds) processes were not shared.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2026, sampling from: air water channel, cfu: 1 cfu, bacterial identification: roseomonas mucosa.Sampling date: (b)(6) 2026, sampling from: instrument channel, cfu: 3 cfu, bacterial identification: peribacillus sp.Sampling date: (b)(6) 2026, sampling from: auxiliary channel, suction channel, cfu: >1 cfu, bacterial identification: n/a.Sampling date: (b)(6) 2026, sampling from: air water channel, auxiliary channel, instrument channel, suction channel cfu: >1 cfu, bacterial identification: n/a.Sampling date: (b)(6) 2026, sampling from: suction channel, cfu: 1 cfu, bacterial identification: rossellomorea sp.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|