• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXACTAMIX; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXACTAMIX; SET, I.V. FLUID TRANSFER Back to Search Results
Model Number H938738
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an exactamix 500 ml eva (ethylene-vinyl acetate) tpn (total parenteral nutrition) bag leaked from the middle port.This was observed during set up/preparation.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h4, h6 (update codes), h11.H4: the lot was manufactured march 18-19, 2024.H11: the actual device and a photo of the sample were provided for evaluation.Visual inspection was performed on the actual sample which identified leakage into the outer pouch.Visual inspection of the photo was performed and the reported leakage was not easily identified.Functional leak testing of the actual sample was performed and a leak was identified from the administration spike port bonding area.The reported condition was verified.The cause of the condition could not be determined; however, likely due to inadequate or lack of cyclohexanone applied to the spike port cap tube when it was inserted into the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
ECM - AVAILMED S.A. DE C.V. - SP021332
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242705355
MDR Report Key24445830
Report Number1416980-2026-00732
Device Sequence Number16043311
Product Code LHI
UDI-Device Identifier00085412477220
UDI-Public(01)00085412477220(10)60547556(17)270228
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K900585
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberH938738
Device Catalogue NumberH938738
Device Lot Number60547556
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer03/11/2026
Initial Report FDA Received Date02/25/2026
Supplement Report FDA Received Date03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-