STAAR SURGICAL COMPANY EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5 13.2 (-10.50/1.5)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6, investigation type 4110: lens work order search, one similar complaint event(s) within associated lots was found.Claim# (b)(4).
 
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault and significant reduction of ica.Lens explanted and replaced with a backup lens.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Cause of the event was not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1191 walker ave
monrovia, CA 91016
MDR Report Key24445836
Report Number2023826-2026-00332
Device Sequence Number10364624
Product Code QCB
UDI-Device Identifier00840311311477
UDI-Public(01)00840311311477(11)250110(17)271231
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 13.2 (-10.50/1.5)
Device Catalogue NumberVTICM5 13.2
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Initial Report FDA Received Date02/25/2026
Date Device Manufactured01/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MSI-PF, LOT# UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-