| Model Number |
H938175 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that twelve (12) exactamix vented, micro-volume inlets had particles; further described as " black particle on the hose area".The particles appeared to be fixed.This was observed during inspection, shortly after receiving the product.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h11: the reported problem and the lot number combination is the same failure already documented in a field action fa-2024-050.The investigation has been performed and identified the causes of the reported particulate matter within the inlet primary packaging inlet components (including within the sterile fluid path tubing) are related to multiple root causes associated with process/procedural controls, materials, environment, and machine/tooling during manufacturing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|