• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXACTAMIX; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXACTAMIX; SET, I.V. FLUID TRANSFER Back to Search Results
Model Number H938175
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that twelve (12) exactamix vented, micro-volume inlets had particles; further described as " black particle on the hose area".The particles appeared to be fixed.This was observed during inspection, shortly after receiving the product.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: h11: the reported problem and the lot number combination is the same failure already documented in a field action fa-2024-050.The investigation has been performed and identified the causes of the reported particulate matter within the inlet primary packaging inlet components (including within the sterile fluid path tubing) are related to multiple root causes associated with process/procedural controls, materials, environment, and machine/tooling during manufacturing.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
ECM - PLASTICOS Y MATERIAS PRIMAS PYMPSA
3609-2 juan de la barrera
parque industrial el alamo
guadalajara, jalisco 44490
MX   44490
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242705355
MDR Report Key24445882
Report Number1416980-2026-00728
Device Sequence Number21425869
Product Code LHI
UDI-Device Identifier00085412475806
UDI-Public(01)00085412475806(10)803745(17)270321
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K002705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 07/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberH938175
Device Catalogue NumberH938175
Device Lot Number803745
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer06/08/2026
Initial Report FDA Received Date02/25/2026
Supplement Report FDA Received Date07/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3002710954-08/09/24-001-
Patient Sequence Number1
Patient SexUnknown
-
-