• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 382537
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
The catheter and needle were inside the patient and the needle retracted on its own and then once the button was clicked to retract the needle it was extremely slow.
 
Additional Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Additional Manufacturer Narrative
A device history record review was completed by our quality engineer team for provided material number 382537 and lot number 5161790.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported issue, and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect similar issues during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSYTE AUTOGUARD
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key24446086
Report Number1710034-2026-00184
Device Sequence Number16063074
Product Code FOZ
UDI-Device Identifier00382903825370
UDI-Public(01)00382903825370
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number382537
Device Lot Number5161790
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2026
Supplement Date Received by Manufacturer02/27/2026
Initial Report FDA Received Date02/25/2026
Supplement Report FDA Received Date03/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-