| Model Number |
420205-16 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical inc.(isi), has not received the da vinci product with an alleged issue to perform failure analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and found to have a frayed grip cable at the yaw pulley.Some wires were broken, but the cable itself was not fully broken.There was no discoloration, corrosion, or contamination observed on the cable that would occur from improper flushing or rinsing during reprocessing.Further inspection found no abnormally sharp or damaged components that could have contributed to the cable breaking.Common causes of frayed - distal instrument grip cables are attributed to damage to the cable through external collisions, during transport and/or storage, or during use or reprocessing.
|
| |
|