| Catalog Number |
515064 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) follow up mdr for device evaluation: one photo was provided to our quality team for investigation.Upon visual inspection, a hair was observed within the expansion chamber, verifying the reported concern.A device history review was completed for lot 2507403.No deviations or non-conformances were identified during manufacturing that could have contributed to this observation.This product is manufactured in a clean room maintained under positive pressure to minimize the introduction of foreign material.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures, no issues were identified during these inspections.All individuals are required to follow proper gowning procedures before entering the room, including hair protection.Although no direct issues were identified, we have identified the root cause is related to personnel failing to follow proper gowning procedures, allowing a hair to enter the product during assembly.Manufacturing personnel have been notified of this incident to increase awareness of this matter.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd protector p20-o had foreign matter.The following information was provided by the initial reporter: material #515064, batch #2507403.Rcc received a complaint via phone.Pir attached.Phaseal optima protector (p20-0) had a black hair sealed inside of the adaptor.Phaseal optima protector was contaminated and could have been used to treat a patient who may have been immunocompromised.Fortunately, this was caught, documented, and the adaptor was discarded.Product#: 515064.
|
| |
|
Search Alerts/Recalls
|