• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL OPTIMA; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL OPTIMA; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM Back to Search Results
Catalog Number 515064
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) follow up mdr for device evaluation: one photo was provided to our quality team for investigation.Upon visual inspection, a hair was observed within the expansion chamber, verifying the reported concern.A device history review was completed for lot 2507403.No deviations or non-conformances were identified during manufacturing that could have contributed to this observation.This product is manufactured in a clean room maintained under positive pressure to minimize the introduction of foreign material.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures, no issues were identified during these inspections.All individuals are required to follow proper gowning procedures before entering the room, including hair protection.Although no direct issues were identified, we have identified the root cause is related to personnel failing to follow proper gowning procedures, allowing a hair to enter the product during assembly.Manufacturing personnel have been notified of this incident to increase awareness of this matter.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd protector p20-o had foreign matter.The following information was provided by the initial reporter: material #515064, batch #2507403.Rcc received a complaint via phone.Pir attached.Phaseal optima protector (p20-0) had a black hair sealed inside of the adaptor.Phaseal optima protector was contaminated and could have been used to treat a patient who may have been immunocompromised.Fortunately, this was caught, documented, and the adaptor was discarded.Product#: 515064.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL OPTIMA
Common Device Name
CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key24449851
Report Number3003152976-2026-00103
Device Sequence Number13274819
Product Code ONB
UDI-Device Identifier00382905150647
UDI-Public(01)00382905150647(10)2507403
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K181221
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number515064
Device Lot Number2507403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2026
Supplement Date Received by Manufacturer04/10/2026
Initial Report FDA Received Date02/25/2026
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-