It was reported that after an initial bunion surgery on (b)(6) 2025, the hardware had loosened and a broken screw was discovered via radiographic image during a post-op follow-up.No other patient outcomes were reported as a result of this event.The device was not returned for evaluation as it currently remains implanted in the patient.Additional information indicates the hardware may potentially be removed at a later date.The patient is currently in normal shoes with arch support and experiences pain when on her feet and when elevating her feet.The device history records for kits used in this case were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.The most likely cause of the reported event could not be determined as the device was not returned for evaluation.However, the device was likely subjected to excessive bending stresses which resulted in its breakage.The company will supplement the mdr as necessary and appropriate.Additional tmc devices explanted in the same revision surgery were reported in.
|
|
It was reported that after an initial bunion surgery on (b)(6) 2025, the hardware had loosened and a broken screw was discovered via radiographic image during a post-op follow-up.No other patient outcomes were reported as a result of this event.All hardware currently remains implanted.Although there has been no injury reported and this malfunction has not resulted in a revision surgery to date, there have been similar events where this malfunction has resulted in a revision surgery.Therefore, an mdr will be filed to ensure full compliance with 21 cfr part 803.
|
|
Additional information received on 03/18/2026.Updated fields are listed below.B1.Adverse event.B2.Outcome attributed to adverse event: required intervention to prevent permanent impairment/damage.B4.Date of this report: 03/31/2026.B5.Describe event or problem: it was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed and replaced during a recent revision surgery due to pain, swelling, loosened and broken hardware.The revision surgery was approximately three weeks prior to (b)(6) 2026, however the exact date is unknown.The hardware had loosened and a broken screw was discovered via radiographic image during a post-op follow-up.The patient reported that she subsequently developed hammer toes of the 2nd, 3rd, and 4th toes, and experienced pain when putting on shoes and if on the foot for more than an hour.No other patient outcomes were reported as a result of this event.D1.Brand name: lapiplasty speedplate.D4.Model #: sk50.Lot #: tm1298.Expiration date (dd-mmm-yyyy): 10-jan-2030.Primary unique device identifier (udi) #: (b)(4).G3.Date received by manufacturer (dd-mmm-yyyy): 18-mar-2026.G4.Premarket identification, pma/510(k): k251135.G6.Type of report: 30-day, follow-up (1).H2.If follow-up, what type? additional information.H4.Device manufacture date (mm/dd/yyyy): 01/10/2025.H6.Adverse event problem.Health effect - clinical code (e): 4577 (swelling/edema).Health effect - impact code (f): 4629 (device revision or replacement).Medical device problem code (a): 2993 (adverse event without identified device or use problem).Component code (g): 4781 (fixation tines).H11.Manufacturer narrative: it was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed and replaced during a recent revision surgery due to pain, swelling, loosened and broken hardware.The revision surgery was approximately three weeks prior to (b)(6) 2026, however the exact date is unknown.The hardware had loosened and a broken screw was discovered via radiographic image during a post-op follow-up.The patient reported that she subsequently developed hammer toes of the 2nd, 3rd, and 4th toes, and experienced pain when putting on shoes and if on the foot for more than an hour.The patient is currently recovering and prescribed six weeks of no weight bearing.No other patient outcomes were reported as a result of this event.No devices were returned for evaluation.The device history records for kits used in this case were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.The most likely cause of the reported event could not be determined based on the information provided and no device being returned.However, the device was likely subjected to excessive bending stresses which resulted in its breakage.The company will supplement the mdr as necessary and appropriate.Additional tmc devices explanted in the same revision surgery were reported in 3011623994-2026-00031, 3011623994-2026-00064, 3011623994-2026-00065, 3011623994-2026-00066, 3011623994-2026-00067.
|
|
It was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed and replaced during a recent revision surgery due to pain, swelling, loosened and broken hardware.The revision surgery was approximately three weeks prior to (b)(6) 2026, however the exact date is unknown.The hardware had loosened and a broken screw was discovered via radiographic image during a post-op follow-up.The patient reported that she subsequently developed hammer toes of the 2nd, 3rd, and 4th toes, and experienced pain when putting on shoes and if on the foot for more than an hour.No other patient outcomes were reported as a result of this event.
|