Model Number 400272-03
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The harmonic ace instrument was analyzed and found to have a broken blade.The blade broke at roughly.233¿ from the base.Broken piece was returned.The probable root cause of the broken blade is attributed to use conditions.Broken blades result from blade scratches and damage caused by contact with other instruments or other hard/metal objects (e.G., staples, clips, etc.) during use while the instrument is activated.Blade fractures propagate from initial contact sites due to the ultrasonic vibration of the harmonic ace blade, leading to eventual/premature failure (e.G., scratches and damage result in cracks, which then result in broken blades).
 
Event or Problem Description
It was reported that during a da vinci-assisted prostatectomy surgical procedure, the harmonic ace broke at the jaw during use.The procedure was completed with no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Common Device Name
HARMONIC ACE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key24450215
Report Number2955842-2026-10797
Device Sequence Number9336616
Product Code NAY
UDI-Device Identifier10886874111083
UDI-Public(01)10886874111083(10)L90210610
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K112584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number400272-03
Device Catalogue Number400272
Device Lot NumberL90210610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/02/2023
Initial Report FDA Received Date02/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES