|
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported complaint.The insert was found with a broken curved blade.Reported complaint.Broken piece, approximately 0.65" x 0.10" was returned.No material appeared to be missing as the broken assembly was pieced back together.The probable root cause of the broken blade is attributed to use conditions.Broken blades result from blade scratches and damage caused by contact with other instruments or other hard/metal objects (e.G., staples, clips, etc.) during use while the instrument is activated.Blade fractures propagate from initial contact sites due to the ultrasonic vibration of the harmonic ace blade, leading to eventual/premature failure (e.G., scratches and damage result in cracks, which then result in broken blades).
|