Device 24 of 30 based on the available information, this event is deemed to be a reportable malfunction.Batch record review: lot 5k03190 was manufactured on 13/oct/2025, in convex 2 pc line, with a total of (b)(4) market units (mkus).The complaints engineer performed a batch record review on 09/feb/2026, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material identification (id) 1161271 and manufacturing order (b)(4).Review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.The reported malfunction could be generated during the manufacturing process of the final line, therefore no further batch record review of the subassembly lots was required.Photograph, video and/or physical sample evaluation: there were photographs associated with this case and in these, the reported defect can be seen.No unused return sample was expected.Historical complaints review: on 09/feb/2026, complaints engineer ran a query in database in order to verify the complaints reported for the 5k03190 lot for the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ defect and no additional complaints were identified.Historical nonconformance review: on 09/feb/2026, complaints engineer ran a query in database looking for any in process nonconformance / corrective and preventive actions (capa) (s) associated to the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ defect for the lot number 5k03190 and as result, no nonconformance / corrective and preventive actions (capa) (s) for this malfunction were generated during the manufacturing process of the referenced lot.Current quality controls: based on the process instruction (pi), the following tests are performed in the manufacturing lines, in order to identify this failure mode in our manufacturing process: adhesive to convex disc centering (visual): not less than (nlt) 0.095" (2.4mm) from id of the adhesive center hole to id of the convex disc at any point.(qc tool) picture frame srp: not less than (nlt) 0.504" (12.8mm), width of the srp left at any point on the picture frame.(qc tool) picture frame srp to convex disc centering: not less than (nlt) 0.040" (1.0mm) from the outer diameter of thermoformed disc to collar srp kiss cut inner diameter.(qc tool) frequency: hourly sample quantity: 5 samples.Defect rate analysis: there have been 30 defective parts confirmed to date from a lot size of (b)(4) products.This represents a defect rate of only (b)(4), which was well within an appropriate acceptable quality level (aql) for this defect which should be 2.5% based on our standard operating procedure (sop).In addition, all of the in-process testing on this lot did not find a single defective unit, which confirms that the lot was unlikely to breach an acceptable quality level (aql) of 2.5.To date, it was well within our accepted acceptable quality level (aql) for this type of failure mode or defect.Conclusion: a review of batch records was completed and showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancies related to this issue were found within the documentation.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
|