Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
|
Product complaint # (b)(4).Investigation summary: according to the information provided: "it was reported that the impacted products had an unknown allegation.It did not happened in surgery" and "there is no lot code on them.The ring that holds the liner fell off.Cannot identify.The instruments sent back were getting complaints from multiple doctors due to the wear and tear over the years".The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found signs of heavy usage and deep scratches on its overall surface.The threaded insert and snap ring were not returned for evaluation.No additional anomaly was identified on the device surface.Scratches observed are consistent with other tools and hard edges coming in contact with the device.However, taking in consideration the aspect and age of the device, the observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As for the components that were not returned for evaluation, condition may had occurred during the cleaning/sterilization process, particularly when the components can be assembled or disassembled.As these procedures are often repeated, there is an increased risk of components going missing.However, without concrete evidence or further information, it is not possible to determine a root cause.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the pin trl lnr neut 600dx36id would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was able to be retrieved for this device.If the lot/serial number becomes available, the record will be re-assessed.Corrected: h3.
|