• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH HYDROSET/10CC INJECTBLE CEMENT; METHYL METHACRYLATE FOR CRANIOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH HYDROSET/10CC INJECTBLE CEMENT; METHYL METHACRYLATE FOR CRANIOPLASTY Back to Search Results
Catalog Number 79-43910
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Convulsion/Seizure (4406)
Date of Event 02/04/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received, it will be reported on a supplemental report.
 
Event or Problem Description
It was reported that the product worked as intended during the case; however, the patient with cement implant experienced convulsions afterward.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
Additional information: h6.Correction: d2a.The reported convulsions occurred following implantation; however, the product was reported to have functioned as intended during the procedure.Additional information and medical assessment identified multiple potential causes of postoperative convulsions unrelated to the device, including metabolic, medication related, infectious, anesthetic, and neurological factors.Review of the ifu confirmed relevant contraindications and patient considerations.Based on the available information, no evidence was identified that the device caused or contributed to the reported event.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSET/10CC INJECTBLE CEMENT
Common Device Name
METHYL METHACRYLATE FOR CRANIOPLASTY
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
Stryker Orthopaedics-Mahwah
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
elena gehring
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
+49()7614888208
MDR Report Key24454835
Report Number0008010177-2026-00018
Device Sequence Number10148566
Product Code GXP
UDI-Device Identifier07613154275014
UDI-Public07613154275014
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Type of Report Initial,Followup
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number79-43910
Device Lot Number4C5S7IC03840
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2026
Supplement Date Received by Manufacturer02/04/2026
Initial Report FDA Received Date02/26/2026
Supplement Report FDA Received Date09/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-