• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA AMISTEM-P; AMISTEM-P LAT. SIZE4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA AMISTEM-P; AMISTEM-P LAT. SIZE4 Back to Search Results
Model Number 01.18.414
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 01/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
At about 21 days after the primary, the patient came in presenting pain due to a periprosthetic fracture and the cause is unknown.The surgeon cabled the fracture and revised the amistem p stem to an m-vizion stem with proximal body.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 02 february 2026.Lot 2300470: (b)(4) items manufactured and released on 31-may-2023.Expiration date: 2028-05-14.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported events during the period of review.Root cause: based on the information available no definitive root cause can be established, while there is no indication that any potential issue with the device may have caused or contributed to the event, and the device history record review does not indicate any potential manufacturing related issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMISTEM-P
Common Device Name
AMISTEM-P LAT. SIZE4
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key24454904
Report Number3005180920-2026-00106
Device Sequence Number13254068
Product Code LZO
UDI-Device Identifier07630040720137
UDI-Public07630040720137
Combination Product (Y/N)N
PMA/510(K) Number
K173794
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.18.414
Device Lot Number2300470
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Initial Report FDA Received Date02/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-