| Model Number |
I-RES WARMER 1.0 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.Legal manufacturer: gehc trout - 3114 n grandview blvd.Usa waukesha, wi 53188.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a malfunction resulting in loss of oxygen therapy.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The issue was resolved by tightening the o2 fitting.
|
| |
|
Search Alerts/Recalls
|