| Catalog Number |
201101210 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during hip replacement surgery, actis reamer tip broke off and was retained in the femoral canal.Reamer was being used to expand the inner diameter of the femur to accommodate the femoral component.The reamer broke at the junction of the reamer leaving approximately 6cm piece within the shaft of the femur.It was determined to be unsafe to try and remove the piece and in order to not compromise the femur it was left in place.Surgery was completed and it was determined that this would not require any changes in postoperative care.Patient discharged in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|