TORNIER S.A.S. AEQUALIS REVERSED GLENOSPHERE ECCENTRIC 42MM X 29MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWD195
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Subluxation (4525)
Date of Event 01/30/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
It was reported that, the patient sadly had an alcohol problem hence the many falls leading to dislocations, not compliant with rehab and more likely to infect.This is the fourth revision of the spacer, liner and glenosphere.No further information was provided.
 
Event or Problem Description
It was reported that, the patient sadly had an alcohol problem hence the many falls leading to dislocations, not compliant with rehab and more likely to infect.This is the fourth revision of the spacer, liner and glenosphere.No further information was provided.
 
Additional Manufacturer Narrative
Correction- added linked mfr numbers of this case.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.A device inspection was not possible since the affected device was not returned for investigation.From the only provided picture taken in the operating room just after the devices explantation soiled implants, only the lot numbers of the insert and glenosphere could be identified.No additional information could be observed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.It is specified in the ifu as follows- the following situations threaten the success of the shoulder replacement implant, people likely to fall, alcoholism or drug abuse, inability of the patient to follow the surgeon's recommendations and the physical therapy program.Based on the investigation, the root cause was attributed to a patient related issue.The event was most likely caused by a patient's failure to follow instructions alcohol problem resulting in numerous falls, dislocations and revisions, with an increased risk of infection.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS REVERSED GLENOSPHERE ECCENTRIC 42MM X 29MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key24458228
Report Number3000931034-2026-00223
Device Sequence Number10148737
Product Code KWS
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081059
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial,Followup
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWD195
Device Lot Number3906BB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2026
Supplement Date Received by Manufacturer04/17/2026
Initial Report FDA Received Date02/26/2026
Supplement Report FDA Received Date05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-