| Model Number |
VICM5_12.1(-3.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Due to refractive surprise, the lens was exchanged for a same model, length but different power length lens.The problem was resolved.Cause of the event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no additional similar complaint type events within associated lots were found.Refractive surprise: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.The lens was exchanged for a same model, length but different power lens.The problem was resolved.Cause of the event is unknown.H3 - device evaluation: the lens was returned dry in a microcentrifuge tube with residue/debris on the lens.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens to be within specifications.(b)(4).
|
| |
|