|
Multiple attempts were made to obtain additional information and receive the device back in-house.The device was not returned to stryker sustainability solutions for evaluation.Therefore, the reported event was not confirmed.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: generator delivers more energy than displayed settings membrane switch assembly damage tissue accumulation between the blade and shaft prolonged activation (in general or against solid surfaces), repeated use of instrument beyond intended use gasket damage allowing for fluid ingress into the shaft assembly user selects improper min generator settings user activates on max instead of min.The instructions for use (ifu) state: minimally invasive procedures should be performed only by persons having adequate training and familiarity with minimally invasive techniques.Consult medical literature relative to techniques, complications, and hazards prior to performance of any minimally invasive procedure.Minimally invasive instruments may vary in diameter from manufacturer to manufacturer.When minimally invasive instruments and accessories from different manufacturers are employed together in a procedure, verify compatibility prior to initiation of the procedure.A thorough understanding of the principles and techniques involved in laser, electrosurgical, and ultrasonic procedures is essential to avoid shock and burn hazards to both patient and medical personnel and damage to the device or other medical instruments.Ensure that electrical insulation or grounding is not compromised.Do not immerse instruments in liquid unless the instruments are designed and labeled to be immersed.Verify compatibility with generators.Use device only with ethicon endo-surgery generator g11 (gen11) software version 2013_1, 2014_1, or 2016_1.Software revision can be found under "system information" in the generator g11 (gen11) "settings" menu.Refer to the generator g11 (gen11) operator's manual for more information.In case of system failure, ensure the availability of the appropriate back up equipment relevant to the specific procedure.Audible high-pitched ringing, resonating from the blade or hand piece, are an abnormal condition and an indicator that the blade or hand piece is not operating properly.The ringing may be an indicator that the hand piece is beyond its useful life or that the blade has not been attached properly, which may result in abnormally high shaft temperatures and user or patient injury.The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the advanced hemostasis hand control button.Do not attempt to seal vessels in excess of 7 mm in diameter.Blood and tissue buildup between the blade and shaft may result in abnormally high temperatures at the distal end of the shaft.To prevent burn injury, remove any visible tissue buildup at the distal end of the shaft.As with all energy sources (electrosurgery, laser, or ultrasound), there are concerns about the carcinogenic and infectious potential of the by-products, such as tissue smoke plume and aerosols.Appropriate measures such as protective eyewear, filtration masks, and effective smoke evacuation equipment should be used in both open and laparoscopic procedures.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.¿ to avoid user or patient injury in the event that accidental activation occurs, the instrument blade, clamp arm, and distal end of the shaft should not be in contact with the patient, drapes, or flammable materials while not in use.During and following activation in tissue, the instrument blade, clamp arm, and distal 7 cm of the shaft may be hot.Avoid unintended contact with tissue, drapes, and surgical gowns at all times.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Do not introduce or withdraw the instrument with the jaws open through a trocar sleeve as this may damage the instrument.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this occurs, there may be an instrument failure, and the generator touchscreen displays a troubleshooting message.To avoid user or patient injury, do not activate an electrosurgical device in close proximity to the harmonic instruments.The aerosols created by the activation of the harmonic instruments in fatty tissue are potentially flammable.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad and increased blade, clamp arm and distal shaft temperatures.The entire exposed blade tip and any exposed blade shaft are active and will cut/coagulate tissue when the instrument blade is activated.Be careful to avoid inadvertent contact between all exposed blade surfaces and surrounding tissue when using the instrument.Use only the appropriate foot switch, hand piece, instruments, and power cord to ensure that they are compatible with the generator.After removing the instrument, examine the tissue for hemostasis.If hemostasis is not present, appropriate techniques should be used to achieve hemostasis.Successful hemostasis may require adjunct measures when harmonic instruments are used on solid organs.Due to the difficulty of visualizing internal structures, proceed slowly and do not attempt to transect large masses of tissue in one activation.Avoid the division of large vascular/biliary bundles when using the instrument under these conditions.If activation is unintentionally stopped while sealing, maintain jaw closure and reactivate.Do not use advanced hemostasis mode for procedures where energy application is desired prior to full closure of the jaws (e.G.Solid organs).Energy is not delivered in advanced hemostasis mode until the jaws are completely closed.During benchtop testing of vessels >5 mm, the strongest vessel seals were achieved by allowing the advanced hemostasis mode to completely transect the targeted vessel.Prolonged usage of advanced hemostasis mode may cause tissue pad damage.Products manufactured or distributed by companies not authorized by ethicon endo-surgery may not be compatible with the harmonic system.Use of such products may lead to unanticipated results and possible injury to the user or patient.If during use the generator displays ''advanced features are not available in this device", the following functions of adaptive tissue technology are no longer available: regulated energy delivery, enhanced audible feedback and advanced hemostasis.As a consequence, the device vessel sealing indication does not exceed 5 mm.Instruments or devices which come into contact with bodily fluids may require special disposal handling to prevent biological contamination.¿ incidental and prolonged activation against solid surfaces, such as bone, may result in blade heating and subsequent blade failure, and should be avoided.Dispose of all opened instruments whether used or unused.This device is packaged and sterilized for single use only.Multiple patient use may compromise the device integrity or create a risk of contamination that, in turn, may result in patient injury or illness.Directions for use 1.Verify compatibility of all instruments and accessories prior to using this instrument (refer to warnings and precautions).2.The hand piece is shipped non-sterile.It must be sterilized prior to each use according to the instructions for use supplied with the hand piece.Assembly 1.Using sterile technique, remove the instrument from the package.To avoid damage, do not flip the instrument into the sterile field.2.While holding the hand piece in a vertical orientation, attach the hand piece to the instrument by rotating the instrument onto the hand piece in a clockwise rotation as viewed from the distal end of the instrument (finger tight only).3.Use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.4.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.5.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.6.Note: hold only the gray hand piece and not the instrument handle while applying the torque wrench (illustration 2).7.Close the trigger.Remove the torque wrench by sliding it off of the shaft.Do not dispose of the torque wrench until the procedure is completed.The torque wrench is used to remove the instrument from the hand piece following the procedure (illustration 3).Dispose of the torque wrench only after completing the procedure.8.Note: take care to avoid damage to the blade and clamp arm by closing the trigger while sliding the torque wrench onto or off of the shaft.Page 7 of 19 reprocessed harmonic ace®+7, 5mm diameter shears with advanced hemostasis 9.Note: take care to avoid injury from the blade tip while sliding the torque wrench onto or off of the shaft.Operation 1.Connect the assembled hand piece and instrument to the generator and turn the generator power on.Note: min is the only mode that can be adjusted.2.Select the desired minimum power level using the increase and decrease "buttons on the generator touchscreen.Note: the recommended minimum starting power level is level 3 (illustration 4).For greater tissue cutting speed use a higher generator power level, and for greater coagulation use a lower generator power level.The amount of energy delivered to the tissue and resultant tissue effects are a function of many factors, including the power level selected, blade characteristics, grip force, tissue tension, tissue type, pathology, and surgical technique.3.For optimal performance and to avoid tissue sticking, clean the instrument blade, clamp arm, and distal end of the shaft throughout the procedure by activating the instrument tip in saline (illustration 5).Note: do not touch the instrument to metal while activated (illustration 6).Note: do not clean the blade tip with abrasives.It can be wiped with a moist gauze sponge to remove tissue, if necessary.If tissue is still visible in the clamp arm, use hemostats to remove residue, taking care not to actuate the hand piece.If desired, the instrument may be unplugged (illustration 7).4.5.6.The blade is ultrasonically energized when either the foot switch pedal is depressed or one of the hand control buttons is depressed.Pressing the left foot pedal of the footswitch or the lower hand control button (min) on the instrument activates the selected minimum power level.Pressing the right foot pedal of the footswitch or upper hand control button (max) on the instrument activates the maximum power level.Pressing the advanced hemostasis button on the instrument activates the advanced hemostasis.Under default settings, advanced hemostasis is hand-activated only.Note: the "min" button can be re-assigned to advanced hemostasis by selecting "convert min to advanced hemostasis" on the gen11 "settings¿ screen.Refer to the generator g11 operator's manual "settings" section (illustration 8).Once "convert min to advanced hemostasis" is selected, the advanced hemostasis button is disabled.Note: energy is not delivered in advanced hemostasis mode until the jaws are completely closed.Caution: if activation is unintentionally stopped while sealing, maintain jaw closure and reactivate.Note: the generator provides an audible tone to indicate when the instrument blade is active.The generator changes to a second activation tone as adaptive tissue technology regulates the delivery of energy.Thermal influences such as fluids or minimal to no tissue in the jaws may affect the presence or timing of the tone change.The tone change does not provide confirmation of tissue effect.When the second tone is heard, the situation should be assessed and the intended surgical action completed, such as gradual application of tension to facilitate transection.The secondary activation tone change is not a substitute for surgical experience.Note: scratches on the blade may lead to premature blade failure.Avoid accidental contact with other instruments during use.Do not use any means other than the torque wrench to attach or detach the instrument from the hand piece.Close the clamp arm by closing the trigger, and insert the shaft through a trocar or incision.The instrument can be used for dissection, grasping, coagulation, and cutting between the blade and clamp arm.Note: to achieve complete sealing, the trigger should be fully closed and the vessel fully contained between clamp arm and blade of device.An audible and tactile "click" indicates full trigger closure.To achieve full closure of the jaws of the device, squeeze the plastic trigger until you feel it stop against the plastic handle (plastic to plastic) (illustration 9).If full trigger closure is released prior to or during activation on tissue, an audible and tactile "click" is evident.Increase grip force until full trigger closure is achieved.Note: keep clamp arm open when using the inside bottom of the blade for back cutting (illustration 10).Warning: do not use advanced hemostasis mode for procedures where energy application is desired prior to full closure of the jaws (e.G.Solid organs).Energy is not delivered in advanced hemostasis mode until the jaws are completely closed.Warning: during benchtop testing of vessels >5 mm, the strongest vessel seals were achieved by allowing the advanced hemostasis mode to completely transect the targeted vessel.Warning: prolonged usage of advanced hemostasis mode may cause tissue pad damage.7.Close the clamp arm by closing the trigger and remove the shaft from the trocar or incision.Shaft rotation 1.The instrument¿s shaft can be rotated 360°to facilitate visualization and access to target tissue when dissecting, grasping, coagulating, and cutting.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|