• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 490037
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: rn complained of ail, with no bubbles visible.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was provided for evaluation.The sample was visually evaluated, and no defects were observed.To attempt replication of the reported defect of air-in-line alarm on the pump, the sample was attached to the pump and then tested by running therapy while staggering the flow rate throughout the therapy.Additionally, a functional leakage test was performed on the sample, and no leakage was detected.A measurement of the outer diameter of the pump segment tubing was taken and found to be 0.152 inches, which meets the specification of 0.152-0.154 inches.Based on the investigation findings, the sample met internal specification; therefore, the reported defect was not confirmed to be due to the manufacturing process.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT®
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
8334251464
MDR Report Key24460649
Report Number2521402-2026-00247
Device Sequence Number15778995
Product Code FPA
UDI-Device Identifier04046964185922
UDI-Public(01)04046964185922(11)250617(17)300531(10)0062011343
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number490037
Device Lot Number0062011343
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date02/26/2026
Supplement Report FDA Received Date03/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-