This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was provided for evaluation.The sample was visually evaluated, and no defects were observed.To attempt replication of the reported defect of air-in-line alarm on the pump, the sample was attached to the pump and then tested by running therapy while staggering the flow rate throughout the therapy.Additionally, a functional leakage test was performed on the sample, and no leakage was detected.A measurement of the outer diameter of the pump segment tubing was taken and found to be 0.152 inches, which meets the specification of 0.152-0.154 inches.Based on the investigation findings, the sample met internal specification; therefore, the reported defect was not confirmed to be due to the manufacturing process.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|