Model Number 8670055
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis #708790239:product: 8670055, lot: it20a002 visual and microscopic examination confirmed that the tip of the instrument the tap was broken.This failure is consistent with overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a tap used for spinal th erapy.It was reported that the tip of the tapper was open and could not be used prior to the tlif (transforaminal lumbar interbody fusion) procedure.The guidewire could not thread through and the product was used before. there was no patient involvement reported.There were no further complications reported regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEXTANT
Common Device Name
TAP, BONE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key24465966
Report Number3003120897-2026-00315
Device Sequence Number13157005
Product Code HWX
UDI-Device Identifier00721902300813
UDI-Public(01)00721902300813(11)210104(10)IT20A002
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8670055
Device Catalogue Number8670055
Device Lot NumberIT20A002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2026
Initial Report FDA Received Date02/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-