ALCON RESEARCH, LLC - HOUSTON UNITY TOTAL PLUS COMBINED PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000484
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
 
Event or Problem Description
A health care professional reported that the tip of an infusion line was clogged with one pak during set up for vitrectomy surgery.The surgery was completed post replacement of console but nothing was wrong with console as it was confirmed that the infusion cannula was found to be clogged.There was no impact to the patient.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The wet pack was visually inspected.Flash in the straight through connector restricted fluid to flow to the infusion cannula tip.The system is designed to perform quality safety to detect for this unwarranted condition and alert the user during priming, prior to surgery.The straight through connector is a molded component by our supplier manufacturer.The investigation conducted a non-conformance review of the reported lot number.No relevant deviations were identified; all corresponding production release specifications defined in the device master record were met.The root cause of the customer¿s complaint is related to insufficient draft on the core pin led to material adhesion and subsequent degradation of the shutoff surface, resulting in intermittent flash and/or occlusion of the molded bore.Action was taken and to address root cause the supplier has replaced damaged core pins, repaired tool damage, and completed gate area tooling repairs.Production of the part will also be run at nominal parameters for a minimum prescribed period with engineering batch review, along with engineering trials to evaluate increased draft effectiveness.Additionally, supplier will execute verification production runs under heightened inspection to confirm absence of stretched necks, flash, or bore occlusion.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNITY TOTAL PLUS COMBINED PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24466911
Report Number1644019-2026-01110
Device Sequence Number18473056
Product Code LRO
UDI-Device Identifier00380650004840
UDI-Public(01)00380650004840(17)270531(11)250605(10)17FFH5
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065000484
Device Lot Number17FFH5
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/05/2026
Supplement Date Received by Manufacturer05/05/2026
Initial Report FDA Received Date02/27/2026
Supplement Report FDA Received Date05/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1