| Model Number |
6893200 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the flow transducer test failed during pre-use check and replacement of the o2 gas module solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The o2 gas module regulates the inspiratory gas flow and a faulty o2 gas module may affect the delivered volume or gas mixture (o2/air).The reported issue was confirmed in provided device's logs.The defective part has not been returned fot the further investigation, despite request.Our conclusion is that the replaced part was the cause of the failure.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator failed pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
|
| |
|