| Model Number |
LUCAS 3 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A stryker service representative performed an evaluation of the customer¿s device and was able to duplicate the reported issue.After replacing protective pcb, control pcb and completing some unrelated repairs, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device alarmed and paused.In this state, a delay in cardiopulmonary resuscitation may occur.If a device malfunctions, the device operating instructions indicate that the user is to remove the device and immediately start manual chest compressions.There was no patient use associated with the reported event.
|
| |
|
Search Alerts/Recalls
|