U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND ATTUNE PS FB INSRT SZ 6 14MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND ATTUNE PS FB INSRT SZ 6 14MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151640614
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Loss of Range of Motion (2032); Discomfort (2330); Joint Laxity (4526)
Date of Event 02/12/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient was playing pickle ball when patient felt discomfort and pain in knee.Patient described knee feeling stuck, but after some time patient was able to move knee normally.Discomfort continued leading patient to see a doctor for knee.Surgeon found that there was some resistance when moving the knee as well as slight laxity.Poly exchange was performed to correct problem.Surgeon found scar tissue within that notch of the ps femur that was catching the ps post within the notch.Surgeon removed scar tissue and trialed ps poly's and determined a thicker poly was now needed due to the laxity of the joint.Doi: unknown.Doe: (b)(6) 2026.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 6 14MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24469350
Report Number1818910-2026-03815
Device Sequence Number10230393
Product Code JWH
UDI-Device Identifier10603295050124
UDI-Public(01)10603295050124(17)170531(10)239599
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number151640614
Device Lot Number239599
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/12/2026
Supplement Date Received by Manufacturer03/06/2026
Initial Report FDA Received Date02/27/2026
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/17/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-