Model Number PXYAT0
Medical Device Problem Codes Failure to Deliver (2338); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nonhealth care professional reported that during the intraocular lens (iol) implantation surgery, the lens did not deploy correctly and plunger rode over the lens.Additional information was received stating that there was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX PRO TRIFOCAL IOL W/AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24469475
Report Number9612169-2026-00439
Device Sequence Number18537082
Product Code HQL
UDI-Device Identifier00380658493271
UDI-Public(01)00380658493271(17)301115(11)251117(10)26159258
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPXYAT0
Device Catalogue NumberPXYAT0.195
Device Lot Number26159258
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2026
Supplement Date Received by Manufacturer05/06/2026
Initial Report FDA Received Date02/27/2026
Supplement Report FDA Received Date05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
-
-