• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ UNILATERAL IMPLANT, OVERSIZED; Joint, Temporomandibular, Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ UNILATERAL IMPLANT, OVERSIZED; Joint, Temporomandibular, Implant Back to Search Results
Catalog Number CHG024
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received, it will be reported on a supplemental report.
 
Event or Problem Description
It was reported that the patient has a bilateral dislocation.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
Additional information: d4, d6a, h4, h6.Correction: d2a.The reported bilateral dislocation was confirmed based on review of the postoperative scans.Clinical and medical assessment determined that the dislocation was related to the patient's pre reconstruction condition, including lack of masticatory muscle control, absence of mandibular support, and lack of occlusion.Review of the device history records confirmed that the devices were manufactured according to specifications.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ UNILATERAL IMPLANT, OVERSIZED
Common Device Name
Joint, Temporomandibular, Implant
Manufacturer (Section D)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer (Section G)
TMJ Solutions, LLC dba TMJ Concepts
6059 king drive
ventura 93003
Manufacturer Contact
elena gehring
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
+49()7614888208
MDR Report Key24469937
Report Number0002031049-2026-00048
Device Sequence Number18719103
Product Code LZD
UDI-Device Identifier07613327626605
UDI-Public07613327626605
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberCHG024
Device Lot Number2507081064
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer02/02/2026
Initial Report FDA Received Date02/27/2026
Supplement Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/06/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-