• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BURR, OVAL, 8 FLUTE 4.0MM X 13CM; BURR,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BURR, OVAL, 8 FLUTE 4.0MM X 13CM; BURR,SURGICAL Back to Search Results
Model Number BURR, OVAL, 8 FLUTE 4.0MM X 13CM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
On 06-feb-2026, an arthrex subsidiary employee reported via (b)(4) that when a new ar-8400obe 8-flute oval burr was attached, it was unstable.When they removed it, the blade attachment broke.They opened another new one, but it felt as if it did not have enough torque, so they gave up on using it.Another replacement was used, and the procedure was completed successfully with no patient harm.Additional information was received on 12-feb-2026.The rep advised that they do not know what procedure was performed.They advised that all broken fragments were retrieved, and no dualwave outflow tubing was used.The rep then stated they did not know if there was an uninterrupted flow of irrigation through the device during its use, and they did not know if any shaver consoles or handpieces were used with the attachment.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 -complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to allowing the rotating portion to touch any metallic object during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, OVAL, 8 FLUTE 4.0MM X 13CM
Common Device Name
BURR,SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 -194
Manufacturer Contact
jared engle
8009337001-73216
MDR Report Key24470167
Report Number1220246-2026-01060
Device Sequence Number10308464
Product Code GFF
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBURR, OVAL, 8 FLUTE 4.0MM X 13CM
Device Catalogue NumberAR-8400OBE
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2026
Supplement Date Received by Manufacturer06/18/2026
Initial Report FDA Received Date02/27/2026
Supplement Report FDA Received Date06/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-