| Model Number |
8.5 MM FLIP CUTTER II |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during the all-inside anterior cruciate ligament surgery, the blade of the flipcutter could no longer be retracted.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.One unpackaged ar-1204af-85, flipcutter®, batch 5217171248, was received for investigation.Visual inspection revealed a fractured tip.Functional testing could not be performed due to the damage to the device.The most likely cause of the reported failure is off-axis insertion or prying and leveraging the device with excessive force.The complaint allegation is confirmed.Refer to the investigation photos.
|
| |
|