| Catalog Number |
00515047501 |
| Medical Device Problem Code |
Unintended Power Up (1162)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).The device was returned sealed in the sterile packaging.Functional testing of the returned product found the device would not power on with the original battery pack but did power on and function normally in both modes when connected to a lab battery pack.Visual inspection of the interior of the pack found the batteries had leaked electrolyte inside of the pack.There did not appear to be damage to the wiring of the pack.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing and/or design issue.The reported event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.E1 telephone number: (b)(6).G2 country: japan.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery the device started working inside of the outer box during transportation.There was no patient impact or delay.During product evaluation it was found that the battery leaked electrolyte.Due diligence is complete and there is no additional information available.
|
| |
|