| Model Number |
IV |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis will be completed.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A health professional reported that during loading, the plunger bent and could not enter the cartridge.The plunger did not contact the intraocular lens (iol).There was no patient contact.Additional information has been requested but is not available at the time of this report.
|
| |
|
Event or Problem Description
|
|
Additional information has been received stated that there was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided b.5.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was not received at the manufacturing site for evaluation for the report of a bent plunger; therefore, the condition of the product could not be verified.Device/batch record review: a review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Conformance achieved: the investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Inconclusive: based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.Potential contributing factors: while the root cause and its origin are inconclusive, the following factors outlined in the reported product¿s instruction for use (ifu) could potentially contribute to an outcome similar to the reported event.The ifu states that the handpiece must be inspected prior to each use to confirm it is free of damage.It further emphasizes that if the handpiece¿particularly the plunger tip¿appears damaged, bent, or otherwise unfit for proper use, it must not be used.In such cases, users are instructed to contact company immediately.The ifu also provides a list of qualified cartridge combinations to use.The use of an unqualified combination may cause damage to the intraocular lens (iol) and potential complication during the implantation process.The manufacturer internal reference number is: (b)(4).
|
| |
|