• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 511.801
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: investigation summary: a visual and functional assessment was performed on the device: the following steps failed: general condition check function in running mode check the oscillation frequency with frequency meter during the service evaluation the following failures were identified: functional: frozen/will not move visual: component damaged conclusion: ¿ failure reported is confirmed ¿ root cause: faulty parts ¿ action: repair.
 
Event or Problem Description
It was reported that during service and evaluation, it was determined that the saw attachment device was frozen/would not move.It was noted in the service order that the device adaptor was blocked causing it not to work at all.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILLATING SAW ATTACHMENT II
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
elba bello
eimattstrasse 3
oberdorf 04436
SZ   04436
9083863534
MDR Report Key24471410
Report Number8030965-2026-02012
Device Sequence Number15856875
Product Code GFA
UDI-Device Identifier07611819102262
UDI-Public(01)07611819102262
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number511.801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2026
Initial Report FDA Received Date02/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-