SYNTHES GMBH OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
| Catalog Number |
511.801 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: investigation summary: a visual and functional assessment was performed on the device: the following steps failed: general condition check function in running mode check the oscillation frequency with frequency meter during the service evaluation the following failures were identified: functional: frozen/will not move visual: component damaged conclusion: ¿ failure reported is confirmed ¿ root cause: faulty parts ¿ action: repair.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the saw attachment device was frozen/would not move.It was noted in the service order that the device adaptor was blocked causing it not to work at all.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|