|
SMITH & NEPHEW, INC. ULTRATAPE 2MM BLUE 38 PKG OF 6; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72203896 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a broström ankle ligament repair following proper implant placement, the ultratape 2mm blue 38 pkg of 6 disengaged despite adherence to the established technique.The microraptor knotless sa peek was left inside the patient.The insertion site was cleared of all debris prior to insertion.The procedure was resumed, after a non-significant delay, using an equivalent s+n back-up.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.An evaluation of the customer provided image was performed and found the device with no visible damage.A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Factors that can contributed to the reported event include an impact event to the device inconsistent with norma use.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for assessment.
|
| |
|
Search Alerts/Recalls
|
|
|