| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A non-healthcare professional reported during intraocular lens (iol) implant procedure, the back haptic did not work, it was stuck on the plunger while priming with no patient contact.The procedure was completed on same day.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.9., h.3., h.6.And h.11.The product was returned for analysis, and the reported complaint could not be observed.Intraocular lens (iol) was returned outside of the device.The device was received loose in the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle.No damage observed.The lens was returned outside of the device.Solution is dried on the iol.No damage observed.The product investigation could not identify the root cause for the reported complaint "back haptic did not work: it was stuck on the plunger" as the lens was returned outside of the device.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|