Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant was found to have an area of missing material on the posterior view, measuring approximately 1.5 cm x 1.8 cm.Additionally an area of siltex cracking was identified on the edges of the missing material.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the texture layer and can eventually progress through the shell of texture devices.Siltex cracking is ultimately a result of high stresses.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: migration.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a patient with mentor memorygel breast implants experienced rupture, capsular contracture baker grade unknown, swelling and implant site fluid collection on the left side along with device migration on the right side.As a result, the patient underwent device removal and replacement.This report is for the right breast prosthesis.
|