• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SURGICAL RELIATACK; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SURGICAL RELIATACK; STAPLE, IMPLANTABLE Back to Search Results
Model Number RELTACK5RDPT
Medical Device Problem Codes Noise, Audible (3273); Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: reltack5rdpt, reload reltack5rdpt reliatack 5deeppurc (lot#: n5c2162y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a laparoscopic abdominal wall incisional hernia surgery, a cartridge replacement for all 29 long tacks included in the box (total quantity) was performed without any problems.All tacks were fired without any misfires, but when an additional cartridge containing five tacks was loaded and replaced, an abnormal cracking sound was heard from the handle.There were cases where the tacks fell out, and the fixation was loose.The fallen tacks have been retrieved outside the body to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was returned with a tack protruding from the end of the reload.Inspection of the reload noted that the timing was disrupted.It was reported that an abnormal cracking sound was heard from the handle.The reported issue could not be confirmed.The most likely cause could not be established from the information available.This issue could be due to excessive manipulation or to improper loading and/or unloading of the sulu.It was also reported that there were cases where the tacks fell out and the fixation was loose.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: for loading, do not attempt to remove the shipping wedge until the reload is locked and loaded onto the instrument.Ensure the device is in reload mode, red toggle button exposed, and the device is in the straight (unarticulated) position.To load the reliatack¿ device select the appropriate reload, then pull and hold the blue lock switch button, located on top of the instrument, in the back position.Insert the pin located at the distal end of the shaft into the reload.Ensure that the arrow on the shaft is aligned with the arrow on the reload.Push the reload onto the distal end of the shaft until the two arrows meet.Release the lock switch button allowing it to return to the forward position.Gently pull on the reload to confirm that the device is properly loaded.Remove the shipping wedge once the reload is properly locked and loaded.Instructions for unloading, push the left side of the toggle button to expose the red portion of the button and ensure that the device is in the straight (unarticulated) position.Pull and hold the blue lock switch button, located on top of the instrument, in the back position.Pull the reload off of the shaft of the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIATACK
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24476433
Report Number1219930-2026-01371
Device Sequence Number18552740
Product Code GDW
UDI-Device Identifier20884521202402
UDI-Public(01)20884521202402(17)270531(10)N4F1724Y
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151659
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRELTACK5RDPT
Device Catalogue NumberRELTACK5RDPT
Device Lot NumberN4F1724Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Supplement Date Received by Manufacturer04/30/2026
Initial Report FDA Received Date02/27/2026
Supplement Report FDA Received Date05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H11
Patient SexUnknown
-
-