| Model Number |
470007-08 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The round tip scissors instrument was analyzed, and the findings did not replicate or confirm the customer reported event.The instrument was tested on an in-house system.The instrument passed the recognition, engagement, and grip tests.The instrument moved intuitively with full range of motion in all directions.The blades opened and closed properly.The instrument was fully functional.No product issue was found.
|
| |
|