| Model Number |
470007-08 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The round tip scissors instrument was analyzed and found to have blade damage at the tip/middle/base of the blade.Both blade edges were indented.The blade indentation caused the cut test failure.The cutting edge did not have corrosion that would have contributed to the blade damage or cutting failure.Common causes of the failure mode mechanical indentations and/or burrs instrument blades are attributed to use-related damage when attempting to cut incompatible material, such as hard objects like bone or cartilage, or from mishandling or misusing the instrument.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|