• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN VALLEYLAB FX8; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN VALLEYLAB FX8; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VLFX8GEN
Medical Device Problem Codes Connection Problem (2900); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Burn(s) (1757); Full thickness (Third Degree) Burn (2696)
Date of Event 02/12/2026
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, the patient had a burn.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: b1, b2, b5, g3, h1, h6 this event has been reassessed and the reportability has been determined to be a reportable serious injury.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter during procedure, the patient had a burn on the buttocks area.Applied prescribed ointments on burn site, a thin layer of antibacterial cream (like silver sulfadiazine) to prevent infection.The burn was caused by a non-medtronic rem pad.A new rem pad was used with the same generator and there was no more burn incident observed.Photo observation showed that the patient had a third degree burn.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the patient had a burn and the burn was caused by a non-medtronic rem pad.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB FX8
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN
6044 longbow court
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN
6044 longbow court
boulder CO 80301 3299
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24479694
Report Number1717344-2026-00316
Device Sequence Number10221394
Product Code GEI
UDI-Device Identifier10884521707306
UDI-Public(01)10884521707306(11)240725(20)02
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K172757
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVLFX8GEN
Device Catalogue NumberVLFX8GEN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Supplement Date Received by Manufacturer03/05/2026
05/03/2026
Initial Report FDA Received Date03/02/2026
Supplement Report FDA Received Date03/19/2026
05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/25/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-