Model Number OPT946
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing f&p's investigation.F&p will provide a follow up report upon completion of f&p's investigation.Product background: the opt946 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan via a fisher and paykel healthcare (f&p) field representative that the tubing of an opt946 optiflow + adult nasal cannula was found detached.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 
9494341005
MDR Report Key24479719
Report Number9611451-2026-00755
Device Sequence Number21396179
Product Code BTT
UDI-Device Identifier09420012429155
UDI-Public(01)09420012429155(10)2103616736(11)250221
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K162553
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT946
Device Catalogue NumberOPT946
Device Lot Number2103616736
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2026
Supplement Date Received by Manufacturer05/15/2026
Initial Report FDA Received Date03/02/2026
Supplement Report FDA Received Date06/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-