| Catalog Number |
71631626 |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6 this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, before a tibial internal fixation, upon opening the sterile packaging of the 3mmx1000mm ball tp gde rd, the sales rep noticed there was a hole in the paper component of the package, rendering the part unsterile and unsuitable for use.The surgery was completed, without any delay, with an equivalent s+n back-up product.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device was not returned for evaluation.However, the photographs were reviewed and revealed a hole in the package.The packaging also appears to be significantly wrinkled.A review made by the quality engineering team revealed that the listed batch use the previous packaging configuration.This product is packaged in a new packaging design that does not cause this issue.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According to the packaging sequence, the device sits in a polyurethane sleeve to protect packaging from damage during transit.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include damage during shipping or mishandling.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|