| Catalog Number |
21-6300-02 |
| Medical Device Problem Codes |
Degraded (1153); Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was stated that the cassette is not detected.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: one device was received for device analysis.Service history review identified the complaint was not related to a previous service of the device within the review period.Functional testing and visual inspection performed during analysis.The customer reported complaint was confirmed.It was found that the downstream occlusion(dso) sensor was worn.The dso sensor was replaced.
|
| |
|