• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PT. CIBA VISION BATAM ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PT. CIBA VISION BATAM ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MA60AC
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Corneal Edema (1791); Endophthalmitis (1835); Hypopyon (1913); Inflammation (1932); Capsular Bag Tear (2639); Toxic Anterior Segment Syndrome (TASS) (4469)
Date of Event 11/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for analysis; the lens remains implanted.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.This report originally filed as (b)(4).The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, patient had developed endophthalmitis.Additional information received stating during a cataract extraction and intraocular lens (iol) implant procedure in the right eye, ophthalmic irrigation and aspiration tip, viscoelastic, and a 3-piece lens were used; the primary incision site clear cornea was 2.4 size, and during surgery a posterior capsular tear occurred with the 3-piece lens, leading to enlargement of the primary incision to 2.75 size with no sutures, wound integrity was intact and negative with no postoperative subconjunctival infection, and postoperative medication drops included antibiotics, steroids, and non-steroidal anti-inflammatory drugs, with it being unknown if the event had resulted in clinically significant visual changes; post-surgery on third day post op less than 14 days later, the patient was diagnosed with endophthalmitis (infectious) and toxic anterior segment syndrome (tass) (2+) with conjunctival inflammation (1+), aqueous cells (2+), and hypopyon, the patient had no pain and the pupil was normal but experienced corneal edema and lid swelling; next day a culture sample was taken and the result was negative, and on the same day the patient was seen by retina and vitrectomy with anterior chamber wash was performed and intracameral (ic) antibiotics were given, the intervention was effective and the current condition of the patient was continuing but improving, with the surgeon¿s opinion being unable to determine if the manufacturer products caused or contributed to the event.This report pertains to one patient from the two patients from the same facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
PT. CIBA VISION BATAM
beringin kav. #204 batamindo
industrial park muka kuning
pulau batam 29433
ID  29433
Manufacturer (Section G)
PT. CIBA VISION BATAM
beringin kav. #204 batamindo
industrial park muka kuning
pulau batam 29433
ID   29433
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24481478
Report Number9681121-2026-00032
Device Sequence Number10230235
Product Code HQL
UDI-Device Identifier00380652552547
UDI-Public(01)00380652552547(17)291209(11)241211(10)31585858
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.210
Device Lot Number31585858
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/06/2026
Initial Report FDA Received Date03/02/2026
Date Device Manufactured12/11/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BETADINE; BETADINE 5%; CENTURION SURGICAL PROCEDURE PAK; CUSTOM-PAK SURGICAL PROCEDURE PACK; LIDOCAINE; MONARCH IOL DELIVERY SYSTEM CARTRIDGE; PHENYLEPHRINE; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE; TETRACAINE; VIGAMOX; VISCOAT OPHTHALMIC VISCOSURGICAL DEVICE
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age74 YR
Patient SexMale
-
-