|
The investigation concludes that a higher-than-expected result was obtained from a single patient sample using vitros ammonia (amon) slide lot: 1020-0268-5934 on a vitros 4600 chemistry system.The assignable cause of the event could not be determined with the information provided.The customer did not provide enough quality control data generated for vitros amon slide lot: 1020-0268-5934 to assess the performance of the slide lot, therefore, a performance issue with vitros amon slide lot: 1020-0268-5934 cannot be ruled out as contributing to the event.An instrument related performance issue did not likely contribute to the event as a vitros amon within-run precision test was within the acceptance criteria.No information was provided concerning the affected sample or how the sample was collected, processed and stored prior to testing.Therefore, it was not possible to establish if the customer is following the sample collection device manufacturer recommendation for sample centrifugation therefore improper pre-analytical sample handling could have contributed to this event.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.
|