• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC MEMORY II DOUBLE LUMEN EXTRACTION BASKET; BILIARY CATHETER AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC MEMORY II DOUBLE LUMEN EXTRACTION BASKET; BILIARY CATHETER AND ACCESSORIES Back to Search Results
Model Number G25149
Medical Device Problem Codes Difficult to Fold, Unfold or Collapse (1254); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with the basket fully retracted.There was no visible sign of damage to the sheath or handle.During our functional testing the device allowed handle manipulation; however, the basket was not responsive to handle manipulation.The handle was disassembled and it was observed that the drive wire was disconnected from the handle with nesting inside the purple hub.For further evaluation of the drive wire cable and basket, the distal catheter was cut to push the basket out of the sheath.The basket was fully formed and intact.Evidence of solder was present on the handle cannula at the joint.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the product said to be involved confirmed the report of unable to manipulate the basket.This report was due to a separation of the drive wire in the device handle.The cause of the separation is unknown.A corrective action (capa) has been initiated to address instances of difficulty during basket advancement and retraction.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action (capa) has been initiated in an effort to reduce occurrences of this nature.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
 
Event or Problem Description
During and endoscopic sphincterotomy for stone extraction in the common bile duct, the physician selected two cook memory ii double lumen extraction baskets.It was initially reported that the assist nurse manipulated the handle to unfold for capturing bile duct small stones, but the handle did not work.There was no reportable information at that time.The device was received on 06feb2026 and the basket would not move when the handle was manipulated.The handle was disassembled, and it was noted the drive wire was disconnected from the handle with nesting in the purple hub.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEMORY II DOUBLE LUMEN EXTRACTION BASKET
Common Device Name
BILIARY CATHETER AND ACCESSORIES
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston salem NC 27105
Manufacturer (Section G)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston salem NC 27105
Manufacturer Contact
sabrina o¿brien
4900 bethania station rd
winston salem, NC 27105
3367440157
MDR Report Key24483100
Report Number1037905-2026-00102
Device Sequence Number21492305
Product Code LQR
UDI-Device Identifier00827002251496
UDI-Public(01)00827002251496(17)281006(10)W4976021
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
K171969
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG25149
Device Catalogue NumberMB-35-2X4-8
Device Lot NumberW4976021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2026
Initial Report FDA Received Date03/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS TJF-260V SCOPE.
Patient SexUnknown
-
-