|
ASCENSION ORTHOPEDICS, INC. BASEPLATE/GLENOSPHERE INSERTER; SHOULDER PROSTHESIS, REVERSE CONFIGURATION
|
Back to Search Results |
|
| Catalog Number |
INS-0990-861 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a reverse total shoulder arthroplasty, while using one (1) baseplate/glenosphere inserter for glenosphere implantation, the tip broke off inside the glenosphere.The surgeon placed the glenosphere onto the morse taper of the baseplate to provisionally engage and tapped the inserter with a mallet on-axis onto the baseplate, and the tip of the inserter broke off inside the glenosphere.They were able to use a free needle to remove the broken tip from the glenosphere and continue the operation.The procedure was resumed, without any delay, using a similar s+n back-up product.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).D4.And h4.Have been updated.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.The visual inspection revealed the device returned heavily worn from use, with the distal threaded tip of the inner shaft fractured off with no material returned.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Additional information: d5, d7a, d9, e1 (name prefix/title), h8.
|
| |
|
Search Alerts/Recalls
|
|
|