Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
|
|
It was reported that during a procedure to repair a fracture around a total hip, the attempt to attach the small gt ring attachment plate to the va-lcp ppfx proximal femur plate failed due to stripped screws on the gt ring attachment.Subsequent attempts to use screws from the large gt ring plate also resulted in stripped screws, affecting all four gt ring plates.The procedure was completed without the gt ring attachment plate, resulting in a delay of approximately 10 minutes.There was no reported patient harm or consequence, and the outcome was good.
|