| Catalog Number |
150611008 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); Insufficient Information (4580)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the attune knee tibial tray was involved in a post-trauma revision of the tibial tray.The event involved a patient who underwent revision surgery after developing a tibial cyst.The clinical history included multiple cementless tibial revisions within a nine-month period, and the surgeon subsequently changed the construct from cementless to hybrid, cementing the tibia for all cases.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional manufacturer narrative: d4: udi: as the catalog/model number was not provided, the (01) gtin is not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Added: b1 product problem corrected: h6, (medical device problem code), (health effect clinical code).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: a1, d4 (expiration date) (lot) (primary udi number), d6a, d6b, d10 (concomitant), g4 (pma/ 510(k)), h4.Corrected: d1, d2a, d2b, d4 (catalog).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received from medical advisor on 14.04.2026.A little more information was provided on the call today by dr.(b)(6).I tried to write down as much as i could, see verbatim notes below.I don't have a clear link on which pc they relate to as i didn't get the patient names.(b)(4)."one lady was pretty big person - solid, she was doing okay and had a fall and kept complaining of knee pain for a long time, 15-19 month showed too much activity on bone scan, she was subsequently revised."just a bit of additional bone scan activity that persisted."electrican went back to work early - after 3m or less the ywere back to work it¿s a physical job in the country".Got the biomechanics reports - fibrosis ingrowth."another retired carpenter - did well for 3 months, jumped into a pool and bottomed out.At the time he almost vomited with pain.Had ongoing discomfort and was revised."another guy who had loosening on femur, retired tough guy, total knee, went on do a 1000km walk 2-3 months post surgery.Suffered aseptic loosening of femur."another lady late 60s, she kept complaining of pain, wondered if she got osteoporosis.Bone scan results showed aseptic loosening of baseplate.Not active lifestyle".
|
| |
|
Search Alerts/Recalls
|
|